Detectors
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 188
11. If necessary, install a new fused silica liner
and Vespel ferrule. When doing so, carefully
feed the fused silica liner through the O-ring
at the top of the transfer line so as not to
damage the O-ring.
12. Carefully replace the fused silica liner, fer-
rule and tube onto the detector base. The
fused silica liner should be positioned so that
it protrudes 3 to 6 mm (1/4-inch) above the
top of the transfer tube weldment. With a
wrench, firmly tighten the transfer tube
1/2-turn past finger-tight.
13. Reinstall the heated block, the heater/sensor
cable assembly, the nut, and the ferrule. The
notch in the bottom of the block fits over the
tubing coming from the detector fitting.
14. Tighten the U-clamp screws, then tighten the
nut on the transfer tube.
15. Place the detector on top of the instrument,
orient it properly, and install the four screws
to hold it. Install the top cover and the
exhaust tubing.
16. Connect the ignitor cable to the printed cir-
cuit board.
17. Install the PMT assembly (or assemblies).
18. Restore normal operating conditions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals