EasyManua.ls Logo

Agilent Technologies 6850 Series - Page 263

Agilent Technologies 6850 Series
337 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
G2629A Control Module
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 264
1. Select Status / Setup / Configure, then
press Edit.
2. Select Edit Oven Type.
Caution
Be sure the GC is physically configured for
the oven type you select. A 200–240 V
instrument will perform abnormally if set to
operate as a 100–120 V oven, while a 100–
120 V oven can be dangerously overheated if
set to operate as a 200–240 V oven.
3. Enter the password as follows: Press dot (. )
then left arrow ( ) to erase the question
mark (?), then enter “. . . (three dots)
as the password. Once this is done, the set-
tings options are enabled.
4. Select the correct oven type, then press OK.

Table of Contents

Related product manuals