Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 285
Case 3: Chromatographic performance
problems suggest a possible leak
This section contains procedures for two differ-
ent inlet system leak tests.
Noninvasive leak check
The first leak test is performed without removing
the customer's column, septum or liner. It is
called the non-invasive leak check. This test does
not prove that the inlet is absolutely leak free,
but is a good first pass or remote diagnostic to
see if complete inlet pressure testing is needed.
1. Confirm that the installed column is config-
ured correctly to the capillary inlet.
2. Put the inlet into splitless mode.
3. Set the purge flow to 50 mL/min; purge time
to 1.0 minute.
4. Set the column flow to 2 mL/min. The
system will try to establish the pneumatic
conditions shown in the following diagram.
The inlet pressure will be a function of the
column dimensions and oven temp. This
example shows 20 psi.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals