EasyManua.ls Logo

Agilent Technologies 6850 Series - Page 299

Agilent Technologies 6850 Series
337 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 300
There will be no carrier flow, because the
column is disconnected and the detector
fitting plugged. Measure the H
2
, air and
makeup gas flows independently by turning
them on one at a time from the GC front
panel or Control Module. They should be
within ±10% of the setpoint. If the flows
are significantly off, the jet could be partially
plugged, there could be a leak anywhere in
the FID pneumatic system, or the detector
EPC module could be defective. Remember
that flow measurement from a soap film
bubble meter are affected by ambient pres-
sure. flow modules are calibrated at Stan-
dard Temperature and Pressure. For accurate
readings, use a true mass flow measurement
device (such as the Agilent Flow Tracker
1000). Resolve any flow control problems
and retest the FID noise.
6. Perform FID maintenance per the GC opera-
tors manual.
Collector
Teflon insulators
Brass castle
assembly
Sealing surface
Partially plugged jet

Table of Contents

Related product manuals