Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 303
Some FID's with very high purity gases and low
bleed columns will have a background under 2 pA
when they are actually lit. If this is the case, the
Lit Offset value can be lowered in the detector
configuration. If the Lit Offset is set too low, the
flame may never try to re-ignite. If the flame is
lit, a cold glass should show condensation if held
above the FID.
To determine if Lit Offset is the cause of the
problem, set it to [0] and confirm that the flame
stays lit for an entire run. If it does, choose a
lower setting for Lit Offset that is appropriate
for the specific FID background of that GC.
Also, the FID temperature should be maintained
at least 20°C > the final oven temp of the GC
analysis. The FID should never be operated at
less than 300°C. This temperature will not
damage columns with a maximum temperature
lower than 300°C.
Procedure
The following procedure will help determine the
root cause(s) for difficulty lighting the FID.
1. Insure the proper jet is installed.
2. Confirm the integrity of the gas supply.
Check gas purities. Leak test the plumbing
modules.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals