EasyManua.ls Logo

Agilent Technologies 6850 Series - Page 311

Agilent Technologies 6850 Series
337 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 312
TCD response and peak shape with respect to
data rate
Areas: C14 C15 C16
These chromatograms were produced on TCD/PPIP - standard checkout sample
20 Hz data rate
85.3 84.2 82.9
Areas: C14 C15 C16
10 Hz data rate
85.8 84.9 83.6
Areas: C14 C15 C16
5 Hz data rate
86.3 85.5 84.1

Table of Contents

Related product manuals