Diagnostics and Troubleshooting
Error Messages
Table 23 No Filament Current Through Deuterium Lamp
Possible Causes Action
Lamp is defective. Replace deuterium lamp.
Electronics failed. Replace spectrophotometer lamp power
supply (SLS) board.
Table 24 Deuterium Lamp Ignition Failed
Possible Causes Action
Lamp is defective. Replace deuterium lamp.
Electronics failed. Replace spectrophotometer lamp power
supply (SLS) board.
Table 25 No Current Through Deuterium Lamp
Possible Causes Action
If diagnostics in your software indicate that
lamp voltage is available, the lamp is
defective.
Replace deuterium lamp.
Electronics failed. Replace spectrophotometer lamp power
supply (SLS) board.