75
Diagnostics and Troubleshooting
Error Messages
Table 26 No Voltage At Deuterium Lamp
Possible Causes Action
Lamp is defective. Replace deuterium lamp.
Electronics failed. Replace spectrophotometer lamp power
supply (SLS) board.
Table 27 No Current Through Tungsten Lamp
Possible Causes Action
If diagnostics in your software indicate that
lamp voltage is available, the lamp is
defective.
Replace tungsten lamp.
Electronics failed. Replace spectrophotometer lamp power
supply (SLS) board.
Table 28 No Voltage At Tungsten Lamp
Possible Causes Action
Lamp is defective. Replace tungsten lamp.
Electronics failed. Replace spectrophotometer lamp power
supply (SLS) board.