91
Diagnostics and Troubleshooting
General Troubleshooting Hints
Measurements are not Reproducible
Causes and solutions for measurements not being reproducible are shown in
Table 47.
Table 47 Measurements are not Reproducible
Possible Causes Action
Very low sample concentration Increase integration time to improve
signal-to-noise ratio.
Variations in the sample Check that the sample does not:
• contain particulates that float in and out
of the light beam (filter if necessary)
• shows signs of bubble formation though
chemical reaction or degassing
• show signs of thermal or photochemical
degradation.
One lamp switched off Check that both lamps are switched on if
measurement over whole spectral range is
required. If spectra in the UV range only are
being measured then deuterium lamp must
be on. If spectra in the Visible range only are
being measured then tungsten lamp must be
on.
Weak lamp Use diagnostics intensity test to check if lamp
intensity is too low and change lamp if
necessary.