Agilent 8453 UV-visible Spectroscopy System Operator’s Manual 51
Good Measurement Practices 3
General Considerations
Blank measurement shows a flat baseline (Figure 9 and Figure 10 on
page 52 show a good and a poor baseline)
Cell orientation of blank and sample measurements is the same
Ideally the cell is not removed between the measurement, which means the
cell is filled/rinsed using a pipette or a flow cell is used
Time between blank and sample measurement should be short
Figure 9 Example of a Blank on Water Showing a Good Baseline