Agilent 8453 UV-visible Spectroscopy System Operator’s Manual 61
Using your Agilent 8453 UV-visible Spectroscopy System 4
Measuring Caffeine Absorbance at 273 nm
6 Press the Blank button on the front of the spectrophotometer or click Blank
on the Instrument Panel to start the measurement.
7 Remember the orientation of the cell in the cell holder. Lift the level to
release the cell, remove it and flush it three times with about 1 ml of your
caffeine sample. Then fill the cell with about 3 ml of the caffeine sample.
Make sure that the cell windows are clean and reposition the cell in the
same orientation as for the reference measurement. Close the cell holder’s
level.
8 Press the Sample button on the front of your spectrophotometer or click
Sample on the Instrument Panel to start the measurement.
9 The view shows you the spectrum of your caffeine sample with a vertical
line indicating your wavelength of interest. Below the spectrum is the
Sample/Result Table which shows the absorbance reading at 273 nm.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals