2 General Power Meter Functions
80 N1911A/1912A P-Series Power Meters User’s Guide
Frequency Sweep Mode
Frequency sweep is generally used in a frequency response calibration
system where the amplitude is fixed, and the frequency of the power
source signal is swept. This mode can be used to determine the frequency
response of a device under test.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a
BNC cable. The same connection is applied for power source TRIG OUT
to power meter TRIG IN. See Figure 2- 58.
3 Press . The Channel Setup displays as shown in Figure 2-59.
NOTE
This feature is only available when N8480 Series, E-Series E4410, E-Series E9300, E-Series
E9320 or P-Series N1920 sensor is connected.
NOTE
When N8480 Series, E-Series E4410 or E-Series E9300 sensor is connected, Sensor Mode
is set to AVG only by default.