5.8 Measuring Optical Amplifier (Wavelength Division Multiplex)
5-77
5
Measurement Functions
Point:
Select the noise level detection position when Detection Type is Point
Auto (Center):
The halfway point to the neighboring signal is used as the
noise level detection position. (See Figure 5.8.3-3) If
there is only one signal, then the noise level dependent on
Resolution during waveform measurement is used as the
detection position. (See Table 5.8.3-2)
Auto (Res): A value dependent on Resolution during waveform
measurement is used as the noise level detection position.
(See Table 5.8.3-2)
Manual: Use a manually set value as the noise level detection
position.
Table 5.8.3-2 Values Dependent on Resolution
Set Resolution (nm) Noise Point (nm)*
0.03
0.50
0.05
0.53
0.07
0.60
0.1
0.65
0.2
0.8
0.5
1.3
1.0
2.2
*: Distance (not span) from the signal wavelength
NF Select
This selects the NF calculation method.
S-ASE: The noise figure is calculated from the beat noise between
the optical signal and the ASE.
Total: The noise figure is calculated from the sum of the
following noise values.
Beat noise between optical signal and ASE
optical output
Beat noise between ASE optical output
Optical signal shot noise
ASE optical output shot noise
Actual Resolution
This selects the Actual Resolution used in the NF calculation.
Measured: This uses a value calculated from a formula. (See 5.8.4
Calculation Method”)
Initial: This uses a value set at time of shipment.