2201R Mobile Tester SCPI Reference Guide
Chapter 8 Configuration commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
144
:CONFigure:EGPRs:BS:MSTAtion:MSLot
Syntax
:CONFigure:EGPRs:BS:MSTAtion:MSLot <int1>
Parameters
int1 is an integer parameter.
The minimum value for int1 is 1, the maximum value is 4. The default value for int1 is 1.
Description
This command either sets the number of UL time slots that the mobile under test is commanded to use during
multislot testing.
Query
The query form of this command returns the current setting. The string delivered back contains one integer.
Example
:CONFigure:EGPRs:BS:MSTAtion:MSLot 4
:CONF:EGPRs:MSTA:MSL?
Value returned in this example: "4".
:CONFigure:EGPRs:BS:BCC
Syntax
:CONFigure:EGPRs:BS:BCC <int1>
Parameters
int1 is an integer.
The minimum value for int1 is 0, the maximum value is 7. The default value for int1 is 0.
Description
This command sets the base station color code.
Query
The query form of this command returns the current setting. The string delivered back contains one integer.
Example
:CONFigure:EGPRs:BS:BCC 1
:CONF:EGPRs:BS:BCC?
Value returned in this example: "1".