EasyManuals Logo

COBHAM 2201R Reference Guide

COBHAM 2201R
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
2201R Mobile Tester SCPI Reference Guide
Chapter 5 System commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
59
:SYSTem:COMMunicate:SERA:REMote
Syntax
:SYSTem:COMMunicate:SERA:REMote <PredefExpr>
Parameters
PredefExpr can take on either of the following expressions: ON|OFF. Default is ON.
Description
Enables or disables remote control of the 2201R via the serial interface port COM 1.
Note: The remote control capability at COM 1 should be disabled prior to controlling external equipment (e.g. the
Bluetooth Connectivity Test products) through this interface.
Query
Returns the current setting.
Example
:SYST:COMM:SERA:REM OFF
:SYST:COMM:SERA:REM?
Returns the following string: "OFF"
:SYSTem:COMMunicate:SERA:TERMinator
Syntax
:SYSTem:COMMunicate:SERA:TERMinator <PredefExpr>
Parameters
PredefExpr can take on either of the following expressions: LF|CR|CRLF.
Description
The command defines the delimiter for SCPI control strings for the RS-232 interface.
Query
Reads and returns the current setting of the terminator used on the RS-232 interface.
Example
:SYST:COMM:SERA:TERM LF
:SYST:COMM:SERA:TERM?
Returns the following string: "LF"