2201R Mobile Tester SCPI Reference Guide
Chapter 8 Configuration commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
161
:CONFigure:EGPRs:BER:RSIGnalling:COUNt
Syntax
:CONFigure:EGPRs:BER:RSIGnalling:COUNt <int1>
Parameters
int1 is an integer parameter.
The minimum value for int1 is 100, the maximum value is 10000. The default value for int1 is 100.
Description
Defines the number of samples (blocks) for each EGPRS reduced signalling BER measurement.
Query
The query form of this command returns the current setting. The string delivered back contains one integer.
Example
:CONF:EGPR:BER:RSIG:COUN 2000
:CONF:EGPR:BER:RSIG:COUN? Value returned: 2000.
:CONFigure:EGPRs:BLER:COUNt
Syntax
:CONFigure:EGPRs:BLER:COUNt <int1>
Parameters
int1 is an integer parameter.
The minimum value for int1 is 0, the maximum value is 10000. The default value for int1 is 100.
Description
Defines the number of samples (blocks) for each BLER measurement. Applies to BLER measurements in EGPRS
synchronous mode.
Query
The query form of this command returns the current setting. The string delivered back contains one integer.
Example
:CONFigure:EGPRs:BLER:COUNt 1
:CONF:EGPRs:BLER:COUNt?
Value returned: "1".