2201R Mobile Tester SCPI Reference Guide
Chapter 10 Measurement commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
207
:MEASure:GSM:ARRay:RFTX:POWer
Syntax
:MEASure:GSM:ARRay:RFTX:POWer <int1>
Parameters
int1 is an integer.
The minimum value for int1 is 0, the maximum value is 1000. The default value for int1 is 0.
Description
Performs the measurement of the mobile RF output power level a specific number of times (set with the int1
parameter). To read out the entire measurement results array, use the :FETCh:GSM:RFTX:POWer command.
Query
The query form of this command performs the measurement the specified number of times (int1 parameter). As soon
as all measurements have been completed, all measurement result values are returned in a string. The string delivered
back contains int1 floating point real numbers with the physical dimension of dBm.
The single measurement results are separated by commas.
Example
:MEASure:GSM:ARRay:RFTX:POWer? 5
The string returned is: "11.22,11.09,11.21,11.14,10.99".
:MEASure:GSM:ARRay:RFTX:TEMPlate
Syntax
:MEASure:GSM:ARRay:RFTX:TEMPlate <int1>
Parameters
int1 is an integer.
The minimum value for int1 is 0, the maximum value is 100. The default value for int1 is 0.
Description
Performs a check of the UL bursts against the power/time template (PTT) a specific number of times (set with the
int1 parameter). To read out the entire measurement results array, use the :FETCh:GSM:RFTX:TEMPlate command.
Note: The power/time template can also be set using the :CALC:RFTX:TEMP commands.
Query
The query form of this command performs the check of the burst shape against the PTT the specified number of
times (int1 parameter). As soon as all checks have been completed, all check result values are returned in a string. The
string delivered back contains int1 boolean numbers. A 0 indicates that there was no violation of the PTT by that
burst, while a 1 means that a burst did violate the PTT.
The single measurement results are separated by commas.
Example
:MEASure:GSM:ARRay:RFTX:TEMPlate 10
:FETCh:GSM:RFTX:TEMP?. The check of the current burst against the PTT is performed ten times. The check result
values are stored internally and are read out by the FETC command. The string returned is: "0,0,0,0,0,0,1,0,0,0",
indicating that the seventh burst checked violated the PTT.