EasyManuals Logo

COBHAM 2201R Reference Guide

COBHAM 2201R
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
2201R Mobile Tester SCPI Reference Guide
Chapter 5 System commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
54
:SYSTem:Message
Syntax
:SYSTem:Message <string1>
Parameters
string1 is a string (text) parameter. The maximum length of string1 is 255 characters.
Description
Writes the message specified with the string1 parameter to the internal system message queue of the instrument.
Query
Reads and returns the oldest unread message in the internal message queue. The string returned contains a maximum
of 255 characters.
Example
:SYSTem:MESSage "23.17,Procedure A5"
:SYST:MESS?
String returned: "23.17,Procedure A5"
In this example, a RAPID! program performs some internal calculations and then writes the result to the system
message queue. This result is then read by the external controller.
:SYSTem:COMMunicate:LOCal
Syntax
:SYSTem:COMMunicate:LOCal
Parameters
There are no parameters.
Description
Sets up the 2201R to allow manual operation on the front panel during SCPI operation.
Note: This command may be used e.g. to allow interactive alignment procedures in a production or quality assurance
flow.
Query
There is no query form of this command available.
Example
:SYSTem:COMM:LOC