ProfiMessage
Theabovegraphicrepresentsthegenerationofmeasurementdatainrelationto
commontriggers.Thetransmissionofthemeasurementdataistimedelayed.
Themax.samplingfrequencyislimitedtothecontinuousmeasurementdatatrans-
missionofmaximumroughly10000samples/sandtothepoweroftheDSP.
Dependingontheuserconfiguration,thefollowingratescanbeachieved:
Transmitted measurement data
No. of parallel or active
channels
Time sig-
nal
1)
FFT/Spectrum
2)
Characteristic
values
3)
8 4 2
X — X
1024
Hz
2048
Hz
4096Hz
— X X
2560
Hz
5120
Hz
10240Hz
— — X
5120
Hz
10240
Hz
20480Hz
1)
Timesignalblock,eachwith1024samples
2)
FFTwith1024samplesor400lines
3)
3characteristicvaluesfromthetimerangeand3characteristicvaluesfromthe
frequencyrange
244
Delphin Technology AG
Dezember 2015
ProfiMessage Manual