EasyManua.ls Logo

Delphin ProfiMessage - 4.6.3.8 Advanced settings tab

Delphin ProfiMessage
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProfiMessage
Nernstfactor=ln10*molargasconstant*absolutetemperature/Faradaycon-
stant
molargasconstant=8.314472inJ/(mol*K),absolutetemperature=temperature
ofthereferencejunction+273.15inK,Faradayconstant=96.4851534inJ/(mV*-
mol)
Sensor type
SelectpH.
Sensor measuring range
Arangeof±625mVissufficientforthissensortype.
Temp. compensation
Temperaturecompensationisdeterminedbymeasuringthetemperatureatthe
measurementpointinthesolutiontobedetermined.Activatecompensationand
configurethechannelusingtheAdvanced settings tab.RefertoSettingsforalldia-
logues.
Scaling
The default scaling uses the standard measurement
methods and should not be altered.
4.6.3.8 Advancedsettingstab
TheExtended settingstabisidenticalforallanalogueinputsensorsbutnotevery
settingisaccessiblefromeverytypeofsensor.
89
Delphin Technology AG
Dezember 2015
ProfiMessage Manual

Table of Contents