ProfiMessage
Analogue inputs
AMDT V1.0 ... 1.2 AMDT V2.0
Measurement
ranges
±10V,alternatively±25Vor
±2.5Vadjustableviajump-
ers
±10V,±5V,±2.5V,±1.25V,
±625mV,±312mV,±156
mV,adjustableviasoftware
Coupling DCcoupling
PresetDC-coupling,choiceof
AC-couplingviaDIPswitch
Max.
samplingfre-
quency
25kHz 20.48kHz
Oversampling No Yes
Anti-aliasing
filter
Setbyhardware(SMDs),pos-
sibletoreplacethesignalcon-
ditioningboard
Automaticallyadapts,with
digitalfilter(DSP)
Potentialisol-
ation
Apotentialdifference
totalling±2.5Vispermitted
betweenthedifferential
inputs('-'connector).
Apotentialdifferencetotalling
upto100Vispermitted
betweenthedifferentialinputs
('-'connector).
Inputpro-
tection
upto25V to±250V
Digital inputs
AMDT V1.0 ... 1.2 AMDT V2.0
Inputhysteresis No Yes
Inputcurrent,type 8mA 3mA
Updating
Coupledtoanalogueinputs(trig-
ger)
Independent
Counter
Frequencymeas-
urement
0.1Hzto100MHz
0.2Hzto30
kHz
Measurementpro-
cedure
Pre-parameterised Automatic
Updating
Coupledtoanalogueinputs(trig-
ger)
Independent