U60 Diagnostic Ultrasound System User Manual Small Parts Measurements and Calculations
- 163 -
2. Measure three pieces of data: L.THY-L, L.THY-W and L.THY-H, by the method of distance
measurement.
Reference Section 6.5.1Generic Measurements in B Mode.
3. After the three measurements, the result of L.THY-V will be displayed in the measurement
result window.
To measure R. THY-V:
1. In the small parts measurement menu, roll the trackball to highlight R.THY-V, press Set.
2. Measure three pieces of data: R.THY-L, R.THY-W and R.THY-H, by the method of distance
measurement.
Reference Section 6.5.1Generic Measurements in B Mode.
3. After the three measurements, the result of R.THY-V will be displayed in the measurement
result window.
To measure Isthmus:
1. In the small parts measurement menu, roll the trackball to highlight Isthmus, press Set.
2. Measure Isthmus by the method of distance measurement.
Reference Section 6.5.1Generic Measurements in B Mode.
3. After the measurement, the result of Isthmus will be displayed in the measurement result
window.
11.2. Small Parts Report
After the small parts examination, the system generates a worksheet.
1. Select the desired probe model and the exam mode
Breast/MSK/Testicle/Superficial/Thyroid.
2. In B mode, after measurements press Report to open Thyroid worksheet in Thyroid exam,
or to open Small Parts worksheet in other small parts exam (such as Breast/ MSK/ Testicle/
Superficial), as shown below: