U60 Diagnostic Ultrasound System User Manual EMC Information
- 282 -
Appendix IV: EMC Information-Guidance and Manufacture’s
Declaration
Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
U60 is intended for use in the electromagnetic environment specified below. The customer or the
user of U60 should assure that they are used in such an environment.
Electromagnetic environment – guidance
U60 uses RF energy only for its internal function.
Therefore, the RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
U60 is suitable for use in all establishments, other
than domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
NOTE:
The EMISSIONS characteristics of U60 make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) U60 might not offer adequate protection to radio-frequency communication services. The
user might need to take mitigation measures, such as relocating or re-orienting the equipment.