U60 Diagnostic Ultrasound System User Manual Specifications
- 192 -
Appendix I: Specifications
A1.1: Electrical Safety Classifications
According to the type of protection against electric
shock
Internally powered equipment,
Class I equipment
According to the degree of protection against
electric shock
Type BF
According to the degree of protection against
harmful ingress of liquid
Whole device: Ordinary equipment (Sealed
equipment without liquid proof);
Probe (not including the probe connector):
IPX7;
Footswitch (optional): IP68.
According to the degree of safety of application in
the presence of a flammable gas
Equipment not suitable for use in the presence
of a flammable gas
According to the mode of operation
Continuous operation
According to the grade of EMC
CISPR 11 Group 1, Class A
Standards Compliance
EN 60601-1:2006:A1+2013
idt IEC 60601-1: 2005:A1+2012
EN 60601-1-2: 2015
idt IEC 60601-1-2: 2014
EN 60601-2-37:2008/A1:2015
idt IEC 60601-2-37:2007/A1:2015
A1.2: Power Supply
Operating Voltage
100 V-240 V~
Operating Frequency
50 Hz/60 Hz
Input Current
1.8A-0.8A
Lithium battery
Capacity
5000 mAh
Voltage
14.8 VDC
Average working time
1 h
Maximum charging time
8 h
Cycle life
300 times