EasyManua.ls Logo

EDAN U60 - A2.8.11. Test of Probe L552 UB

EDAN U60
302 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
U60 Diagnostic Ultrasound System User Manual Ultrasound Intensity and Safety
- 274 -
A2.8.11. Test of Probe L552UB
Acoustic Output Reporting Table for IEC60601-2-37
Operating Mode: B mode
Transducer: L552UB
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.71
0.60
0.60
N/A
Index Component Value
0.60
0.60
0.60
0.60
Associated
acoustic
parameters
p
r.
α
at zMI
(MPa)
1.53
P
(mW)
26.58
26.58
N/A
P1X1
(mW)
26.58
26.58
zs
(cm)
2.00
zb
(cm)
2.00
zMI
(cm)
2.05
zPII,a
(cm)
2.05
fawf
(MHz)
4.71
4.74
4.74
4.74
4.74
N/A
Other
Information
prr
(Hz)
-
Srr
(Hz)
110.98
npps
1.00
I
pi.
α
at zPII,a
(W/cm
2
)
157.30
Ispta,a at zPII,a or zSII,a
(mW/cm
2
)
20.45
Ispta at zPII or zSII
(mW/cm
2
)
44.23
pr at zPII
(MPa)
2.25
Operating
control
conditions
Focus Depth(mm)
25
25
25
25
25
N/A
Display Depth(mm)
29
29
29
29
29
N/A
Frequency Index
4.5
4.5
4.5
4.5
4.5
N/A
ScanAngle
level 0
level 0
level 0
level 0
level 0
N/A