U60 Diagnostic Ultrasound System User Manual System Control
- 61 -
2. Roll the trackball to highlight a user-defined comment, and press Set.
3. Press Del to delete the comment.
4. Press OK to save the modification, or press Cancel to give up and close the dialog box.
5.7.9. Preset Data
Press Set up, roll the trackball to highlight Preset Data, and you can view three options in Preset
Data submenu: Import, Export and Factory Default.
To Export Preset Data
This function is used to save all user preset data to an USB disk for back-up.
Roll the trackball to highlight Export and press Set. The system displays a dialog
"Import/export will cover the files that have the same file names! Contiunes? ". Click "OK"
to continue. The preset data named "Settings" will be exported to the U50 folder in the USB
disk..
To Import Preset Data
This function is used to import the existing preset data to the preset data memory of the
system. The system will be reset and operate according to the newly imported preset data.
Roll the trackball to highlight Import and press Set. The system displays a dialog "Preset
data exists, replace it?". Click "OK" to continue. The preset data file should be named
"Settings" and put under the U50 folder of the USB disk.
WARNING
Be careful, if you confirm to import existing preset data, all of the previous preset will be
cleared, and the system will operate according to the newly imported preset data.
To Restore Factory Default Presetting
Roll the trackball to highlight Factory Default and press Set. The system will restore the
factory presetting.
WARNING
Be careful, if you confirm to restore factory default presetting, all of the previous preset
will be cleared, and the system will restore preset data at the factory.