U60 Diagnostic Ultrasound System User Manual Inspection and Maintenance
- 182 -
15.2.1.4. Storage
WARNING
1. Dry the transducer after high-level disinfection or sterilization and store it in sterile
environment.
2. Do not use the carrying case for storing the transducer, because the carrying case
may become a source of infection.
1. Ensure the transducer is cleaned, disinfected, sterilized and completely dried before
storage.
2. Store the transducer in a sterile environment or in a disposable sterile package.
3. Store the transducer under the following conditions:
a) Atmospheric Temp.: -20℃〜+55℃
b) Relative Humidity: 15%~95% (Non-condensing)
c) Atmospheric Pressure: 70kPa ~ 106kPa.
15.2.2. Needle Guide Bracket Cleaning and Sterilization
NOTE:
1. Use proper sterilization technique at all times when performing a biopsy.
2. Ensure that protective gloves are worn.
WARNING
1. The needle guide bracket kits are not disinfected or sterilized before delivery. The
operators must clean and sterilize the needle guide bracket kits before the first use
and after each subsequent use.
2. Inspect the bracket for damage such as cracks or breakage. If damage is evident,
discontinue use of bracket and contact your Edan representative for disposal
guidance.
3. Sterilize the bracket before disposal or sending back to manufacturer for repair.