U60 Diagnostic Ultrasound System User Manual Inspection and Maintenance
- 178 -
Transducer Applied Region
Transducer Type
Disinfecting Intensity
Disinfecting Method
Contact intact body surface
Body surface
LLD
Spraying or wiping
Contact mucous membrane
Intra-cavity
HLD
Immersion
Note:
LLD=Low-level Disinfection; HLD=High-level disinfection
The validated disinfectants for transducer are:
Disinfectants
Disinfecting Intensity
Disinfecting Method
Ethanol (75%)
LLD
Spraying or wiping
Isopropanol (70%)
LLD
Spraying or wiping
Cidex OPA (0.55%)
HLD
Immersion
Cidex Glutaraldehyde(2.4%)
HLD
Immersion
WARNING
1. Unplug the transducer from the system prior to cleaning or disinfecting.
2. To avoid infection, always use protective gloves when performing cleaning and
disinfecting procedures.
3. To avoid infection, ensure that the expiration date of the disinfecting solution has not
passed.
4. Please cleaning the transducer prior to disinfection.
Disinfecting by spraying or wiping:
1. Disconnect the transducer from the system.
2. Wear protective gloves to prevent infection.
3. Clean and dry the transducer according to the methods in section 15.2.1.1 Cleaning.
4. Prepare the disinfectant solution (75% ethanol or 70% isopropanol).
5. Spray the solution to the transducer surface or wipe it with a sterile cloth dampened with the
disinfectant solution. Follow the disinfectant manufacturer's recommended contact time and
mode.
6. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry
sterile cloth or leave the transducer to air dry.
7. Inspect the transducer to ensure that there is no damage.