U60 Diagnostic Ultrasound System User Manual Ultrasound Intensity and Safety
- 200 -
all sources of measurement
uncertainty, based on all
sources being independent.
the hydrophone
calibration certificate.
oscilloscope calibration
certificate.
Table 2 The intensity and pressure measurement uncertainties table
Transducer
Temp
Spatial Ave.
Distortion
Intensity
Pressure
Intensity
Pressure
Intensity
Pressure
C5-2b
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
L15-7b
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
P5-1b
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
C352UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
L742UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
L1042UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
E612UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
C422UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
C612UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
C6152UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
L552UB
±2.40%
±1.20%
±17.5%
±8.75%
±4%
±2%
Notes
Uncertainties based on
temperature variation of
the water bath and
difference in the mean
temperature between the
hydrophone calibration
lab and our facility.
Uncertainties based
on part 7 of A2.7
and discussions with
industry consultants.
Non-linear distortion
estimates are based on part 7
of A2.7 and discussions with
industry consultants.
Table 3 The power, center frequency and MI measurement uncertainties table
Transducer
Total uncertainties
Power
Center frequency
MI
C5-2b
±22.91%
±0.2%
±11.68%
L15-7b
±22.91%
±0.2%
±11.68%
P5-1b
±22.91%
±0.2%
±11.68%
C352UB
±22.91%
±0.2%
±11.68%
L742UB
±22.91%
±0.2%
±11.68%
L1042UB
±22.91%
±0.2%
±11.68%
E612UB
±22.91%
±0.2%
±11.68%
C422UB
±22.91%
±0.2%
±11.68%
C612UB
±22.91%
±0.2%
±11.68%