U60 Diagnostic Ultrasound System User Manual EMC Information
- 284 -
Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
U60 is intended for use in the electromagnetic environment specified below. The customer or the
user of U60 should assure that they are used in such an environment.
Electromagnetic environment -
guidance
Conducted RF
IEC/EN
61000-4-6
Radiated RF
IEC/EN
61000-4-3
3 V
rms
150 kHz to 80 MHz
6Vrms
c
in ISM bands
between
0,15 MHz and 80
MHz
3 V/m
80 MHz to 2.7 GHz
See table 1
3 V
rms
150 kHz to 80
MHz
6Vrms
c
in ISM
bands
between
0,15 MHz and
80 MHz
3 V/m
80 MHz to 2.7
GHz
Comply with
table 1
Portable and mobile RF communications
equipment should be used no closer to
any part of U60, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
800 MHz to 2.7 GHz
at RF wireless
communications equipment bands
(Portable RF communications
equipment (including peripherals such
as antenna cables and external
antennas) should be used no closer than
30 cm (12 inches) to any part of the U60,
including cables specified by the
manufacturer).
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter