U60 Diagnostic Ultrasound System User Manual System Control
- 43 -
NOTE:
1. The HPRF mode is possible with C5-2b, P5-1b and L15-7b probes only.
2. The HPRF mode can be activated only in real-time mode.
3. The HPRF mode is available according to the following table. At each sample
volume depth, HPRF will first occur at the scale range and PRF listed. Other
higher PRFs and ranges are available up to the highest listing in the table (e.g. for
the C5-2b, at 270mm PRFs 9-12 would be available).
Depth (mm)
C5-2b
P5-1b
L15-7b
70
160 cm/sec PRF=15
80
160 cm/sec PRF=15
90
100 cm/sec PRF=14
160 cm/sec PRF=15
100
100 cm/sec PRF=14
150
N/A (Depth can only reach
100mm)
160
200 cm/sec PRF=12
200 cm/sec PRF=12
N/A
170
160 cm/sec PRF=11
160 cm/sec PRF=11
N/A
180
160 cm/sec PRF=11
160 cm/sec PRF=11
N/A
190
160 cm/sec PRF=11
160 cm/sec PRF=11
N/A
200
160 cm/sec PRF=11
140 cm/sec PRF=10
N/A
210
140 cm/sec PRF=10
140 cm/sec PRF=10
N/A
220
140 cm/sec PRF=10
120 cm/sec PRF=9
N/A
230
120 cm/sec PRF=9
120 cm/sec PRF=9
N/A
240
120 cm/sec PRF=9
120 cm/sec PRF=9
N/A
250
120 cm/sec PRF=9
120 cm/sec PRF=9
N/A
260
120 cm/sec PRF=9
100 cm/sec PRF=8
N/A
270
120 cm/sec PRF=9
100 cm/sec PRF=8
N/A
280
120 cm/sec PRF=9
100 cm/sec PRF=8
N/A
290
100 cm/sec PRF=8
100 cm/sec PRF=8
N/A
300
100 cm/sec PRF=8
100 cm/sec PRF=8
N/A