54200
Users Manual
4-68
This test pattern will be shown differently on a normal or a double scan rate TV receiver.
In general, the double scan rate TV will display a stable picture on the screen when using
this test signal, while a normal TV set will show a very unstable picture with line and
image flicker. Depending on the processing technique of the double scan rate TV
receiver the result may differ a little.
Progressive Scan Check 2
This B/W pattern with color burst offers four vertical white lines on a black background
moving in horizontal direction from left to right. The motion can be switched off at any
position of the screen by selecting the
DIGI_SCAN2_S
test pattern. This pattern contains
50 Hz and 25 Hz test signals for 625 line systems and 60 and 30 Hz signals for 525 line
systems.
The first white line of 200 ns is only present in one field. This will result in a 25 Hz or
30 Hz test signal. The second 200 ns vertical line is visible in the first and second field
producing a normal 50 Hz/ 60 Hz signal. The third line has a width of 370 ns in one field
and 200 ns in the next field. The fourth 370 ns vertical line is identical in both fields.
A. 0.2 µs, one field only
B. 0.2 µs, both fields
C. 0.37 µs one field, 0.2 µs next field
D. 0.37 µs both fields
2 µs distance between each vertical line A to D
Figure 4-102. Progressive Scan Check 2
Applications: See Progressive Scan Check 1.
Progressive Scan Check 3
This B/W pattern with color burst offers five groups of four white lines on a black
background alternating in +45°/-45° orientation. The pattern moves in horizontal
direction from left to right. The motion can be switched off at any position on the screen
by selecting the
DIGI_SCAN3_S
test pattern.
The contents of the four white lines are identical to pattern Progressive Scan Check 2.
A. 0.2 µs, one field only
B. 0.2 µs, both fields
C. 0.37 µs one field, 0.2 µs next field
D. 0.37 µs both fields
2 µs distance between each vertical line A to D
Figure 4-103. Progressive Scan Check 3
Applications: See Progressive Scan Check 1.