Fluke 54200 - Display Mode; Miscellaneous Commands

Fluke 54200
386 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
Remote Control Commands
6
6-39
Miscellaneous commands
Display Mode
Header:
DISP
MODE
Data element:
ON | OF
F
| M
AIN
| SO
UND
| V
IDEO
| ST
ANDARD
| D
IGDATA
|
SU
PPORT
Remark: This command switches the display of the generator on or off (in
remote operation only) and selects the displayed screen. The support
screen can only be shown for information purposes; the settings on this
screen (except the clock) cannot be changed using remote control.
Example:
DISPMODE OFF
or
DISP OF
switched the display of the generator off.
Real-time clock
Header:
TIME
Data element:
String
Remark: The string contains the date for the year, month, day, hour, and minute:
"YYYY.MM.DD,HH:MM"
Range for the date: From January 1st, 1991 to December 31st, 2090,
Range for the time: From 00:00 to 23:59
Example:
TIME "1996.12.24,18:00"
sets the clock to 6.00 p.m. at Christmas Eve 1996.
Local Time Offset
Header:
LTO
FFSET
Data element:
NR1
Remark:
NR1
is the range is from -12 to +12.
The command sets a local time offset related to selected time in hours.
Example:
LTOFFSET 1
sets a time offset of one hour,
LTO 0
sets no time offset.