Table 6-4. Conversion Table
Common Commands and Queries according to IEEE-488.2
Function Command for PM 5415/18 Command for 54200
Clear Status None *CLS
Standard Event Status Enable None *ESE
Operation Complete None *OPC
Recall REC 1...9 *RCL 1...99
Reset None *RST
Save STO 1...9 *SAV 1...99
Service Request Enable MSR *SRE
Wait to Continue None *WAI
Standard Event Status Enable Query None *ESE?
Standard Event Status Register Query None *ESR?
Identification Query ID? *IDN?
Operation Complete Query None *OPC?
Option Query None *OPT?
Service Request Enable Query None *SRE?
Read Status Byte Query None *STB?
Self-test Query TSI *TST?
Learn Device Setup Query IS? *LRN?
Vision Carrier Setting
Function Command for PM 5415/18 Command for 54200
Vision carrier frequency FRQ <NRf>
[Hz/Khz/Mhz]
FREQ <NRf>
[HZ/KHZ/MHZ/GHZ/K/M/G]
Vision carrier level None LEVEl <NRf>
[DBUV,Mv,Uv,V]