EasyManua.ls Logo

Fluke 54200 - General Specifications; Environmental Conditions

Fluke 54200
386 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
General Specifications
7
7-21
General Specifications
Environmental Conditions
Environment Laboratory equipment Class 5 (MIL-T-28800D)
Warming-up time 30 min
Temperature +22 to +24 °C for reference temperature
0 to +50
°
C for operating
-20 to +71
°
C for non-operating
Humidity Relative humidity (MIL-T-28800D)
Operating (no condensation)
+0
°
C to 10
°
C not controlled
+11
°
C to 30
°
C 95%
±
5%
+31
°
C to 40
°
C 75%
±
5%
+41
°
C to 50
°
C 45%
±
5%
Non-operating 5% to 95%
Altitude (MIL-T-28800D)
Operating 3,050 m (10,000 ft)
Non-operating 4,570 m (15,000 ft)
Vibration (MIL-T-28800D)
Operating 0.33 mm
p-p
at 5 Hz to 55 Hz (2 g at 55 Hz)
Non-operating 0.70 mm
p-p
at 10 Hz to 55 Hz,
5 g at 55 Hz to 150 Hz
Shock (MIL-T-28800D)
Operating,
Functional, 18 shocks/20 g
Transport 1 corner / 6 surfaces,
drop height 0.67 m
Safety EN 61010-1 +/A2, Class I
IEC 1010-1 +A1 +A2, Class I
CAN/CSA-C22.2 No 1010.1, Class I
EMC emission EN 55011, Group 1, Class B
VDE 0875, Part 11, Group 1, Class B
CISPR 11, Group 1, Class B
FCC Part 15J Class A
EMC immunity EN 50082-1 (Generic)
RF EM field IEC 801-3/84 (27-500 MHz, 3V/m)
Electrostatic discharge IEC 801-2/84 (
±
8 kV air discharge)
Fast transient IEC 801-4/88 (0.5 kV signal/ control lines, 1 kV mains)
Additional:
RF EM field ENV 50140 / 93 (80-1000 MHz, 3V/m)
Electrostatic discharge IEC 1000-4-2 / 95 (
±
8 kV air discharge),
(
±
4 kV contact discharge)
Magnetic Field IEC 1000-4-8 /93 (50 Hz, 3 A/m)
Recalibration interval 12 months
Reliability MTBF = 20,000 hours

Table of Contents