EasyManua.ls Logo

Fluke 54200 - Patterns

Fluke 54200
386 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54200
Users Manual
7-8
Phase tolerance (PAL/NTSC) 2
°
,
1
°
at reference temperature
Freq. deviation (SECAM)
f
OB
(75% blue) +230 kHz
Tolerance 4 kHz
f
OR
(75% red) -280 kHz
Tolerance 5 kHz
Level 100% (nominal setting)
Tolerance 5% (PAL/NTSC)
10% (SECAM)
2% (PAL/NTSC) at reference temperature for
nominal setting
Setting range 0% to 150%
Tolerance of setting 1 step or 5%, whichever is greater
Resolution 1%
Patterns
Reference ITU Rec. 471-1/1994 and
SMPTE EG27-1994 for Color Bar
SMPTE EG1-1990 for SMPTE Color Bar
CCIR Rec. 473-5,1990 and
CCIR Rec. R26-1981 for IRS 17
CCIR Rec. 473-5,1990 for Multiburst
CCIR Rep. 1221 for PLUGE
Aspect ratio 4:3, 16:9
Circle Centered circle with 4 additional corner circles in 16:9 mode
Center Cross With border castellations
(overscan indication selectable between 2% or 3%)
White 0, 5, 15 to 100% (5% steps) for 625 line systems
7.5, 15 to 100 IRE (5 IRE steps) for 525 line systems
Purity Red, green, blue, cyan, magenta, yellow, white, black
(100/0/75/0 for 625 line systems and
100/7.5/75/7.5 for 525 line systems)
Dots 17x13 dots in 4:3 mode,
23x13 dots in 16:9 mode
With center indication
Crosshatch 18x14 lines in 4:3 mode,
24x14 lines in 16:9 mode
With center indication,
selectable top-left indication and chroma

Table of Contents