GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
3-42 Section 3-6 - System Configuration
3-6-7 System Options Tab
1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
2) Trackball to the System Options tab and press Select. The System Options tab is displayed, as
shown below:
Figure 3-26 System Options Tab
Note: If an option is not available for the specific ultrasound unit, the option name and Key field
is disabled (grey).
3) Trackball to the Key field next to the name of the option being installed, and use the alphanumeric
keyboard to enter the password supplied by GE.
Note: If an incorrect password is entered, if the option has already been correctly installed in the
system, it will be removed.
Note: Please note that all passwords should be visible underneath the control console or/and on
the green password (option) certificate. If a new password (option) is required, contact the
GE representative to initiate the appropriate process, so that the relevant password (option)
will be installed on the system.
4) Trackball to Apply and press Select. The status of the option changes to Installed.
5) Repeat steps 3 and 4 as required.
6) Trackball to the OK button and press Select.