GE Vivid 4 BT03 RFI - CFM Stationary Noise Test; Probe Operation Check (Alignment and Sensitivity) Test

GE Vivid 4 BT03 RFI
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 4 - Functional Checks 4-13
4-4-1-5 CFM Stationary Noise Test
4-4-1-6 Probe Operation Check (Alignment and Sensitivity) Test
Table 4-22 CFM Stationary Noise Test (3S)
Step Task Expected Result(s)
1
Use the vertical nail area in the Electronic
Flow Phantom RMI1425A.
2
Place the 3S probe above the vertical nails
line (position 1).
3
Select a Cardiac preset and activate CFM.
4
Set the following CFM parameters:
Gain: Max possible without color noise
Tissue Priority: 0
•Flash: 0
Ensure that the color does not appear constantly on the white nail.
Table 4-23 Probe Operation Check (Alignment and Sensitivity) Test (3S)
Step Task Expected Result(s)
1
Use the Electronic Flow Phantom RMI1425A.
2
Set the Phantom settings as follows:
Preset: 3
Mode: Steady