GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-23
5.) Unscrew and remove each of the 4 screws that fasten the Monitor Rear Cover in position (two each
side; one upper and one lower), as shown in Figure 8-21.
6.) Carefully lift the Monitor Rear Cover upwards and backwards to remove it from the monitor.
8-3-1-4 Vivid™ 4 17" Monitor Rear Cover Installation
1.) Make sure the Vivid™ 4 control console is lowered to its minimum height.
2.) Carefully lift the replacement Monitor Rear Cover into position on the monitor.
3.) Return and fasten each of the 4 screws (previously removed) to secure the Monitor Rear Cover
firmly (two each side; one upper and one lower), as shown in Figure 8-21 on page 8-22.
4.) Return the two screw covers (previously removed) to the top of the Monitor Rear Cover (one each
side) and place in the closed position - refer to Figure 8-21.
5.) Reconnect the AC power cable to the mains wall outlet.
6.) Turn ON the Vivid™ 4 main circuit breaker, located at the rear of the machine (lower end).