GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-64 Section 8-5 - Back End Parts Replacement
8-5-9 Frame Grabber Board Replacement Procedure
NOTE: The procedures described in this section are only applicable to Vivid™ 4 systems with the
RFI configuration.
8-5-9-1 Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the frame
grabber board replacement procedure.
8-5-9-2 Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
8-5-9-3 Frame Grabber Board Removal Procedure
1) Remove the Vivid™ 4 left and right side covers, and rear cover - see the Rear Cover Removal
Procedure on page 8 - 7 and the Side Covers Installation Procedure on page 8 - 4.
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP side covers, as described in the BEP2 - Side Cover Removal Procedure on page
8 - 54.
4) If applicable, remove the octopus card holder as described in the BEP2 - Octopus Card Holder
Removal Procedure on page 8 - 56.
5) Disconnect the frame grabber cable from the rear of the frame grabber board.
Note: Counting from the top, the frame grabber board is positioned in the fifth PCI slot.
6) Release the securing screw located on the left of the board bracket.
7) Gently remove the frame grabber board from the BEP.
8-5-9-4 Frame Grabber Board Installation Procedure
1) Gently insert the frame grabber board into its slot.
Note: Counting from the top, the frame grabber board is positioned in the fifth PCI slot.
2) Fasten the board with the screw on the left of the bracket.
3) Reconnect the frame grabber cable to the rear of the frame grabber board.
4.) If applicable, install the octopus card holder as described in the BEP2 - Octopus Card Holder
Installation Procedure on page 8 - 56.
5) Reconnect the power cable to the rear of the BEP.
6) Install the BEP side covers, as described in the BEP2 - Side Cover Installation Procedure on page
8 - 54.
7) Return the Vivid™ 4 rear, left and right covers - see the Side Covers Installation Procedure on page
8 - 4 and Rear Cover Replacement Procedures on page 8 - 7.
Note: Perform a complete diagnostic check of the system, as described in
Chapter 7 - Diagnostics/Troubleshooting.