GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
7-4 Section 7-2 - Diagnostics
7-2-3-1 Diagnostic Menu Options
The Diagnostic menu includes the following options:
Software Version: Displays the Vivid™ 4 ultrasound unit software version number and last
upgrade date.
: Displays test results, as described in the Common Service Interface section, on
page 7-86.
: Calibrates the Front End crate, and enables you to perform the display and video
grabbing test, as described in Calibration on page 6 - 11.
: Tests each board in the Front End crate including the TR power supply, the LVPS
supply and temperatures, as described in the Performing Front End (FE) Diagnostics section, on
page 7-5.
: Tests various back end hardware components, as described in the Performing
Back End Diagnostics on the System section, on page 7-69.
: Monitors all low voltages, Tx voltages, temperatures and AC power input.
: Performs a full sequence of Front End tests that do not require user intervention.
Auto Sequence is particularly recommended for detecting problems that are intermittent in nature,
since a test may be paused at any stage and any detected failures reviewed. This is described in
the Auto Sequence section, on page 7-6.
Description Area: Displays a description of the function performed when each of the buttons is
selected. Trackball the cursor over a button to display a description in the Description area.
Report Button: Displays a current diagnostics report.
Exit Button: Closes the Diagnostic menu.