GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-74 Section 8-5 - Back End Parts Replacement
7.) Remove the PC-VIC.
8-5-14-4 PC-VIC Installation Procedure
1) Place a new PC-VIC in the PC-VIC module holder.
2) Return and tighten the four screws (two each side) on the side of the holder - refer to
Figure 8-57, above.
3.) Reconnect all the cables at the rear of the PC-VIC module and the hard disk.
4.) Carefully slide the PC-VIC into position in the same location as that of the removed PC-VIC.
5.) Return and tighten the screw that holds the PC-VIC in position (as shown in Figure 8-56 on page 8-73).
6) Return the BEP side cover as described in the BEP2 - Side Cover Installation Procedure on page
8 - 54.
7) Return the Vivid™ 4 left cover, as described in the Side Covers Installation Procedure on page 8 - 4.
Figure 8-57 PC-VIC Removal