Chapter 8 - Replacement Procedures 8-127
6.) Verify that the switches on the VCR rear panel are as shown in Figure 8-120 below. If necessary,
adjust accordingly.
8-8-3-4 Sony VCR Installation Procedure
1.) Place the VCR on the left side of the Vivid™ 4 ultrasound unit’s top surface (below the control
console - as shown in Figure 8-121 below) and next to the B/W Printer, if present, allowing sufficient
room to access the cables at the rear of the VCR.
Figure 8-121 Video Cassette Recorder (VCR) Installed in the Vivid™ 4 Ultrasound Unit
Figure 8-120 Sony VCR - Switches Setting (Rear Panel)
Switch Position VCR AUDIO SWITCHES (on Rear)
1 OFF Audio/Video Mute during search
2 OFF Search at high speed using FF/REW
3 OFF Reset control counter when cassette is ejected
4 OFF Enable RS232 remote
5 ON Baud rate (see below)
6 ON Baud rate (see below)
123456
OFF
Switch 5 Switch 6 Baud rate
OFF OFF 2400
OFF ON 4800
ON OFF 9500
ON ON 10200
Video Cassette
Recorder (VCR)