GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-15
8-2-8-4 Control Console Upper Cover (Front) Installation Procedure
1) Raise the control console to its maximum height.
2) Return the upper cover (front section) to the Vivid™ 4 control console, making sure it is placed in
the correct position (with screw holes properly aligned).
3.) Turn the metal securing plates underneath the rear of the upper front cover (refer to Figure 8-15),
to fasten the front cover to the upper rear cover and tighten the 2 screws.
4) Return and fasten the 12 screws (6 on each side and 2 in the front as shown in Figure 8-15) to
secure the upper front cover in place.
5.) Raise the front and rear sections of the control console lower cover up into position and tighten the
screws to fasten the lower cover securely. For details refer to the Control Console Lower Cover
Installation Procedure on page 8 - 18.
6.) Return the control console lower rear cover, as described in the Control Console Lower Rear Cover
Installation Procedure on page 8 - 19.