GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
5-54 Section 5-6 - External Peripherals
5-9-3-3 Access / Security
The service interface has different access and security user levels. Users are only granted access to
the tools they are authorized to use.
A modem on the scanner is specifically intended for iLinq use (by the user) and for InSite use.
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
See Chapter 7 - Diagnostics/Troubleshooting for more info about InSite and the Global Service User
Interface.
Table 5-9 Access Authorization
User Level Access Authorization
Operator These service tools are normally used in-house and are protected with a password.
Administrator Use the OnLine Center access method provided by iLinq.
External Service Access is protected with a password.
GE Service Requires knowledge of a password.