IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-65
7-3-9-6-5 Temp Test
The Temp test displays the temperature at the Front End crate, as well as the temperature of the AC
distribution box.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in the Accessing the MUX
Diagnostic Test Options section, on page 7-48.
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in Figure 7-34 on page 7-59.
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
in Figure 7-36 on page 7-61.
4) Trackball to the Temperature button and press Select. The temperatures are displayed, as shown
below:
Figure 7-40 Temperatures
5) Trackball to the Show button and press Select to display the temperatures.
Note: For repeat readings, select Loop every and set the reading time cycle.