GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-125
8-8-2-5 Mitsubishi VCR Removal Procedure
1.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
2.) If optional shelf (P/N H45011CN) is present, refer to the Sony UP-21MD Color Video Printer Removal
Procedure on page 8 - 138.
3.) Disconnect the RS-232C cable, the Audio and S-Video IN and OUT cables, and Power cord from
VCR’s rear as shown in Figure 8-107 on page 8-119.
4.) Remove the VCR from the left side of the Vivid™ 4 ultrasound unit (taking it out from below the
control console and from beside the B/W printer, if present) - refer to Figure 8-106 on page 8-119.