IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-134 Section 8-8 - Peripherals
8-8-4-5 Panasonic VCR Removal Procedure
1.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal
Procedure on page 8 - 8.
2.) Release the B/W printer by undoing the fastening belt around it (together with the VCR, if present).
3.) Disconnect the RS-232C cable (with adaptor), the Audio and S-Video IN and OUT cables, and the
Power cable from the rear of the VCR as shown in Figure 8-122 on page 8-128.
4.) Remove the VCR from the left side of theVivid™ 4 ultrasound unit (taking it out from below the
control console and from beside the B/W printer, if present) - refer to Figure 8-121 on page 8-127.