8-72 Section 8-5 - Back End Parts Replacement
8-5-13-4 BEP2 - ECG Module Installation Procedure
NOTE: BEP2 refers to a Pentium 4-based BEP (newer type).
1.) Slide the new ECG Module into the front of the BEP, pushing it in until it aligns with other modules
at the front of BEP.
2) Secure the ECG Module with four screws, two from each side of the BEP (see Figure 8-54, above).
3.) Replace the BEP Front Panel and secure it with the two screws previously removed.
4) Reconnect the power cable and the COM 2 cable to the rear of the new ECG Module.
5) Install the BEP left and right side covers, as described in the BEP2 - Side Cover Installation
Procedure on page 8 - 54.
6.) Carefully return the BEP to its original location and secure all brackets by tightening the screws.
7) Reconnect all cables at the rear of the BEP.
8) Return the Vivid™ 4 rear cover to its original position - see Rear Cover Replacement Procedures
on page 8 - 7.
9) Return the Vivid™ 4 side covers, as described in the Side Covers Installation Procedure on page 8 - 4.
10) Reconnect the mains AC power cable from the wall outlet.
11) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
12) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
Figure 8-55 ECG Module inside BEP2