EasyManua.ls Logo

GE Vivid 4 BT03 RFI - CFM Noise Floor Test; Penetration Test

GE Vivid 4 BT03 RFI
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
4-12 Section 4-4 - Image Testing: 2D/M/CFM/Doppler
4-4-1-3 2D Penetration Test
4-4-1-4 CFM Noise Floor Test
Table 4-20 2D Noise Penetration Test (3S)
Step Task Expected Result(s)
1
Use the Standard Imaging Phantom
RMI403GS.
2
Select a cardiac preset.
3
Set the following parameters, and scan the
phantom at position (3):
Gain: 37
Power: 0 dB
Depth: 20cm
Focus: Max depth
Dynamic Range: 65dB
4
Record the maximum depth at which tissue
can be differentiated from the noise.
The depth should be greater than 18cm.
Table 4-21 CFM Noise Floor Test (3S)
Step Task Expected Result(s)
1
With the 3S probe in the air, select a Cardiac
preset and activate CFM.
2
Set the CFM ROI to its maximum size.
3
Set the following CFM parameters:
Range: 20cm
Tissue Priority: 0
4
Set the Active Gain until a few color noise dots
appear in the ROI.
The Active Gain should be between 61and 65.

Table of Contents

Related product manuals